Fluke 741B Documenting Process Calibrator Repair
Common issues:
- Won't power on
- LCD display blinks/flashes
- Black lines across LCD display
- Calibration Constant Out Of Bounds error, cannot calibrate
- Won't measure/source mA (milliamp) or volts
- Won't measure/source ohms, TC (thermocouple), or RTD resistance
- Volt/mA/ohms readings unstable or inaccurate
- Can't change date/time
Typical Repair Cost Range: $450 - $775*
* Based on previous repairs; actual cost will be quoted after unit evaluation and may be more or less
Tips & Pointers
If the calibrator will not power on with a known good battery, try powering the unit from the charger adapter alone. NOTE: the proper charger plug is center negative at ~7.5VDC.
If the LCD display blinks or flashes once when the power button is pressed but the unit does not turn on, there is likely a problem with the digital processor and/or the related circuitry. Service will be required.
An unstable or erratic source output for ohms resistance, mA (milliamps), or voltage may indicate an internal precision reference is defective and requires repair.
A "Calibration Constant Out Of Bounds" error indicates an incorrect input was applied during calibration or there is a problem with the internal circuitry. If the correct input levels were applied and the error persists, repair is required.
How To Request Repair
- Tell us what is wrong with the instrument and who to contact by doing one of the following:
- Print our Service Request Form - OR -
- Fill out the repair form below.
- Ship the unit to us using your choice of carrier. Our shipping address depends on what model you are sending and will be shown on the confirmation e-mail.
- We will diagnose the problem and send you a formal quote for repair.
Repair Request Form
Unit Information
Model*
Serial Number
Problem Description
* If your instrument's model is not listed, please note it above and CONTACT US for support options.
Contact Information
Country
Name
Company
Address
Apt, suite, etc.
City
State/Province
Postal Code
Phone 1
Phone 2
E-Mail
SUBMIT
Additional Information
What's next?
After submitting the form above, you will receive an e-mail confirmation with shipping instructions. Ship your unit to our receiving address:
ScopemeterRepair.com, LLC
5641 Broyles Rd
Hartville, MO 65667
United States
ScopemeterRepair.com, LLC
5641 Broyles Rd
Hartville, MO 65667
United States
Payment
No payment is due until all services are completed and the unit is ready to ship. We accept all major credit cards on our website and over the phone, PayPal, and checks or money orders.
Turnaround Time
Turnaround time is typically 5-10 business days but is workload-dependent and may vary. If you require immediate service, CONTACT US to let us know your requirements. We also offer 24-, 48-, and 72-hour turnaround times for most models for an additional fee.